    EDK 6900P-Cl portable Cl2 analyzer

    Overview 

     

    Chlor-alkali process is an industrial process for electrolysis of sodium chloride. It is used in the production of chlorine and sodium hydroxide (alkali/caustic soda), both of which are essential and commonly used chemicals in industry.

     

    Etg6900p-cl was used to analyze Cl2 in hydrogen based on nondispersive ultraviolet absorption spectroscopy (NDUV).  NDUV nondispersive ultraviolet absorption spectroscopy is a common method to determine the concentration of complex gas mixtures. NDUV nondispersive ultraviolet absorption spectroscopy is a common method to determine the concentration of complex gas mixtures. The gas analyzer uses modern optical elements to optimize the results of the analyzer. We used a special LED technology to get a steady emission of radiation. This leads to a high-resolution gas analysis system that can detect changes in gas concentrations below 1ppm.

     

    Nondispersive ultraviolet absorption spectrometry (NDUV) uses the principle that the electrons of a particular gas molecule absorb radiation and then undergo a stimulated transition. Compared with the infrared absorption method, these excitations are affected by a larger cross-section (the absorption coefficient), so the proportion of radiation absorption will be larger. Therefore, NDUV method is particularly suitable for PPM trace gas analysis. In addition, the gas analysis method is not affected by the concentration of interfering water vapor.

     

    Features 

    high performance-price ratio

     "hot" measurement

     Built-in sampling pump     

    Touchable screen

    Zero& span calibration

      Portable rugged shell

    Unique NDUV technology  

     Removable modular construction

     Installation type: plug and play  

    Ethernet or USB remote communication

    Complete heating line and sampling probe

    Rapid Cl2 detection T90 (200ms)

    Modbus, Profibus, Ethernet output


    Portable sampling probe and heating line  

    Installable in flue

    Built-in heating and filtering elements

    flue temperature up to 600 ℃

    Maximum ash load 2g/m3

    Teflon core, dn4/6"

    Integrated portable probe and heating lines

    Manually replaceable filter elements(no tools required)

    Materials: silicone, PTFE, stainless steel, polyamide, etc

     

    ETG 6900 software  

    The program can achieve data storage, chart display, analysis report, real-time detection, data export and other functions; The origin of the time axis can be adjusted to coincide with the start time of measurement, or automatically adapt to the change of the horizontal and vertical axes. The user can see the real-time update of the analysis data in the window. Any existing alerts with reference code are also displayed on the same screen as the software.

     

     

    Specification 

    parameter

    Value/ measurement range

    Gas

    Cl2

    Detection principle

    NDUV  nondispersive ultraviolet absorption spectroscopy

    Range

     Cl20-1000ppm 5% maximum , customizable  H2 0-50%

    Accuracy

    ±2% F.S (Temp. &Press.)

    Precision

    Cl2 0.8ppm

    Operation temp.

      0-50℃

    Zero drift (over 2h)

    Within precision

    Span drift (over 2h)

    Within precision

    Temp. drift

    0.1% reading/℃

    Maximum sample volume

      5(1)L/ min; calibration 3L/min

    Linearity/repeatability

    Within precision