• Photoacoustic Gas Analyzer
  • Dissolved O2, O3, H2 analyzer
  • Sonic densitometer
  • Dissolved carbon dioxide analy
  • Other instrumentation
  • Tracer Gas Monitoring System for Ventilation  DKG-VEN

     Tracer Gas Monitoring System for Ventilation

     DKG-VEN

    Introduction

    The time spent indoor is increasing with the modernization and urbanization process, making the indoor air quality IAQ)more important than ever.

    IAQ relies on many factors, such as temperature, humidity, the concentration of CO2 and VOC’s. One of the effective measures to adjust and improve IAQ is ventilation system.

    One of the methods to evaluate ventilation, including natural ventilation and mechanical ventilation, is to monitor the concentration change of tracer gas to calculate the ventilation efficiency. The mostly used tracer gas is selected from SF6, freon 134a, N2O, ethane or CO2.

    Features

    Photoacoustic Spectroscopy with patented cantilever pressure sensor

    Low detection limithigh accuracy

    Fast response time

    To measure tracer gas such SF6, Freon 134a, N2O,ethane, CO2 and VOC’s

    Multipoint sampler, extended 12 to 64 sampling point

    No pre-treatment, No consumables, No carrier gas 

    Applications

    Ventilation evaluation in buildings like air exchange, air of age

    Ventilation efficiency for HVAC

    Ventilation for fume hood


    Air flow in duct

    Pollution spread research

    Particle, virus, toxic gas stimulation

    Ventilation test for SEMI S6

    Ventilation and leak test in nuclear powerplant

    Data for CFD research

    Specifications

    Gas Monitoring——Tracer gas SF6,Freon134a,N2O,CO2,ethane and VOC’s,300+ gases can be selected

    Measurement range

    0.1 ppb-10 ppm

    Dynamic rate

    1:100000

    Detection limit

    0.1 ppb

    Accuracy

    2%

    Response time

    3-10s

    Calibration

    Factory calibration done before delivery

    Operation Environment

    Temperature

    0℃ ~ +45℃

    Humidity

    Below 90% RH, no condense

    Pressure

    Ambient pressure

    Enclosure

    IP20 (IEC 529)

    Vibration test

    33 Hz strong vibration

    Storage Environment

    Temperature

    -20℃ ~ +60℃

    Sampling Environment

    Temperature

    0 ~ +49℃ no condense

    Pressure

    930 mbar ~ 1100 mbar

    Humidity

    Tdew +8 ℃ or higher

    Flow rate

    60 ml/min ~ 300 ml/min

    Particle

    < 1 μm  

    Dimension

    48.4cm Wx13.9cm Hx40.5cm D (19.1in Wx5.5in Hx16.0in D)

    Weight

    Appx. 13 kg

    Connector

    Specified quick plug for 3mm diameter tube,various plugs can be ordered upon request

    Electric parameter

    Power rating

    100 ~ 240VAC, 50 ~ 60 Hz

      Power consumption

    100W

    Interface

    Ethernet、USB、AK

    The sampling points equipped with monitor, fine particle filter equipped

    Pump in-built, max. 100m sampling distance, PC embedded with 7" front display.

    Data storage appx. 2GB,up to 2-year data (@sampling continuously interval 15s)

    Data transfer via USB disk, USB port, Ethernet or Com port

    Compensate the interference from water vapor, particle automatically

    To work with multipoint sampler to extend to 12/24/32/64 channels

    Low adsorption SiO2 coated stainless steel valve set is used in multipoint sampler

    Working time for each channel can be defined 1~99min in multipoint sampler

    Dosing MFC and other accessories to be ordered upon request


    Consulting phone 010-8243-3533